[:Input]:Current:extsensor:pojump<X; [:Input]:Current:sratio:element1<X - Teledyne T3PM1100 User Manual

Digital power meter
Hide thumbs Also See for T3PM1100:
Table of Contents

Advertisement

Find Quality Products Online at:
Example
:INPUT:CURRENT:EXTSENSOR:CONFIG1 2,0.5,0.1
:INPUT:CURRENT:EXTSENSOR:CONFIG1?
->:INPUT:CURRENT:EXTSENSOR:CONFIG1
2.00E+00,500.0E-03,100.0E-03
[:INPut]:CURRent:EXTSensor:POJump<x>
Sets or returns the jump destination range that is
Description
used when a current peak over-range occurs.
Syntax
[:INPut]:CURRent:EXTSensor:POJump<x>
{OFF|<Voltage>}
Query Syntax
[:INPut]:CURRent:EXTSensor:POJump<x>?
Parameter/
<x>
Return parameter
OFF
<Voltage>
Example
:INPUT:CURRENT:EXTSENSOR:POJUMP1 2V
:INPUT:CURRENT:EXTSENSOR:POJUMP1?
->:INPUT:CURRENT:EXTSENSOR:POJUMP1
2.00E+00
[:INPut]:CURRent:SRATio:ELEMent1<x>
Sets or returns the external current sensor
Description
conversion ratio of the specified element.
Syntax
[:INPut]:CURRent:SRATio:ELEMent1<x> {<NRf>}
Query Syntax
[:INPut]:CURRent:SRATio:ELEMent1<x>?
Parameter/
<x>
Return parameter
<NRf>
152
GlobalTestSupply
www.
T3PM1100 User Manual
Set
Query
1,2(EXT1,EXT2), If <x> is omitted, by
default sets or returns EXT2 config.
No jump destination current range.
See(:INPut:CURRent:RANGe).
Set
Query
1,2(EXT1,EXT2), If <x> is omitted, by
default sets or returns EXT2 config.
0.001 to 9999
.com