Harmonic Measurement

Item
Specifications
Measured item
Voltage, Current, Power
Measured method
Zero-cross simultaneous calculation method
Frequency range
10 Hz to 1.2 kHz.
4096 (Frequency must be 50Hz/60Hz and Update
FFT data length
Rate must be greater than or equal to 0.25S)
Fundamental
Sample rate, window
Frequency
width, and upper limit of
Analysis orders*
45 Hz to 55 Hz
54 Hz to 66 Hz
FFT data length
1024
Fundamental
Frequency
Sample rate, window
10 Hz to 67 Hz
width, and upper limit of
67 Hz to 150 Hz
Analysis orders*
150 Hz to 300 Hz
300 Hz to 600 Hz
600 Hz to 1200 Hz f × 64
Frequency
10 Hz ≤ f < 45 Hz
45 Hz ≤ f < 440 Hz
Accuracy
440 Hz ≤ f <
1.2kHz
GlobalTestSupply
www.
APPENDIX
upper limit
Sample
Window
of Analysis
rate
Width
orders
f × 512
10
50
f x 512
12
50
upper limit
Sample
Window
of Analysis
rate
Width
orders
f × 1024 1
50
f × 512
2
32
f × 256
4
16
f × 128
8
8
16
4
Voltage
Current
Power
0.15% of
0.15% of
0.35% of
reading
reading
reading
+ 0.35%
+ 0.35%
+ 0.50% of
of range
of range
range
0.15% of
0.15% of
0.25% of
reading
reading
reading
+ 0.35%
+ 0.35%
+ 0.50% of
of range
of range
range
0.20% of
0.20% of
0.40% of
reading
reading
reading
+ 0.35%
+ 0.35%
+ 0.50% of
of range
of range
range
207