Aoutput:digital:mode; Aoutput:digital:output - Teledyne T3PM1100 User Manual

Digital power meter
Hide thumbs Also See for T3PM1100:
Table of Contents

Advertisement

Find Quality Products Online at:
Example
:AOUTPUT:RATE1 100,-100
:AOUTPUT:RATE1?
->:AOUTPUT:NORMAL:RATE1 100.0E+00,-100.0E+00
• When the D/A output is in manual range mode
Note
Set the rated value for +5 V output and then that
for –5 V output.
• When the D/A output is in fixed range mode
There is no need to set these values. (The values
do not affect the output operation.)
• When the D/A output is in comparator mode
Set the upper limit and then the lower limit.

:AOUTput:DIGital:MODE

Sets the application mode of digital I/O (Remote
Description
Control Only).
Syntax
:AOUTput:DIGital:MODE {IO|4094}
Query Syntax
:AOUTput:DIGital:MODE?
Parameter/
IO
Return parameter
4094
Example
:AOUTPUT:DIGITAL:MODE IO
:AOUTPUT:DIGITAL:MODE?
->:AOUTPUT:DIGITAL:MODE IO

:AOUTput:DIGital:OUTPut

When the 4094 mode (serial to parallel) is selected
Description
for digital I/O, make use of this command to set
output status.
Syntax
:AOUTput:DIGital:OUTPut {<NR1>,<Boolean>}
GlobalTestSupply
www.
COMMAND OVERVIEW
Set
Query
Sets the digital I/O to IO mode.
Sets the digital I/O to 4094 (serial to
parallel) mode.
Set
.com
131