Advertisement

Navigating From the User Interface

While working from the User Interface, there are four modes of operation that
you can access to view or change system information, parameters, or variables.
While working from the User Interface, the following rules apply:
3-4 | 241CE II Hydrocarbon Dewpoint Analyzer
In RUN mode and Stat\History mode, system information and parameters
can only be viewed (they cannot be changed).
After locating a main menu item, press  repeatedly to cycle through the
parameters associated with that menu.
The software will revert back to the RUN menu if no buttons are pressed
within 2 minutes.
Displayed values can be negative or positive. If the value is negative, '-' will
precede the value.
In ApplicCFG or MonitorCFG mode, parameters or variables can be
viewed and/or changed (passwords are required to enter these modes).
While working from the ApplicCFG or MonitorCFG mode, you can make
changes to as many parameters as you like before exiting the screen.
While working in the ApplicCFG and MonitorCFG menus, use the '+' but-
ton to scroll through the various options for each sub-menu item and to
select an option.
All value and selection changes under the ApplicCFG and MonitorCFG
menus do not take effect until  is pressed to move to the next item or
 is pressed to move to a different main menu level. If configuration
changes are not saved to the onboard EEPROM, the analyzer will revert to
the original settings following a system reset or power-up.
All configuration parameter changes which need to be permanent can
be saved to the onboard EEPROM. To permanently save any configuration
changes, press  to scroll to the "Save AppCFG?" (or "Save AppCFG?")
prompt and then press the '+' button to answer Yes.
PN 903-8797, Rev A