Alarm Led Status - Ametek 241CE II User Manual

Hydrocarbon dewpoint analyzer
Hide thumbs Also See for 241CE II:
Table of Contents

Advertisement

Alarm LED Status

The LEDs on the User Interface are numbered 1–7, from left to right. A lit LED
indicates that the condition associated with the LED currently exists. Four (4)
of the seven (7) LEDs indicate the status of corresponding alarms set by four
alarm relay contacts (two LEDs and their associated alarm relay contacts are
not currently used). The alarm setpoint values for the HCDP temperature are
preset at the factory, but can be changed in the field.
The conditions which define the Warning and Fault alarms are part of the
built-in, continuous diagnostics function. The HCDP temperature alarms will
be set at the end of the Cooling Stage, if the conditions that trigger the alarm
are present. The remaining alarms are set at the end of the stage where the
problem occurred.
The LEDs and their corresponding conditions when turned On are:
PN 903-8797, Rev A
LED #
Setpoint By
Condition Existing When LED Turned On
1
User
The Hydrocarbon Dewpoint temperature Off-Spec Alarm is
disabled (Stat\History : En\Dis Alrm?).
2
Diagnostics
Fault Alarm: Analyzer in Standby Mode – Immediate Service
Required (review SCode and XSCode in the Stat\History
menu to gather information before contacting AMETEK). The
alphanumeric display will also flash to alert the operator to this
condition.
3
Diagnostics
Warning Alarm: Analyzer reporting an abnormal condition –
Investigation Required (review SCode and XSCode in the Stat\
History menu and attempt to correct the problems).
4
Diagnostics
Hydrocarbon Dewpoint temperature Not Found (NoD), or
Purging/Hold Stage Is In Progress (Pur). This setting is made in
the ApplicCFG : UsrRelay menu.
See also the UsrRelay description under "Working in the
Application Configuration Menu" in Chapter 3.
Reserved for Water Content OffSpec Alarm setpoint. Currently
5
User
Not Used.
6
User
The current Hydrocarbon Dewpoint temperature is higher
than the Hydrocarbon Dewpoint temperature Off-Spec Alarm
setpoint value (ApplicCFG : HCDPA). This alarm will only be
seen if the Off-Spec Alarm is enabled in the Stat\History
menu (LED 1 is Off when the Off-Spec Alarm is enabled).
7
User
Reserved for Water Dewpoint temperature OffSpec Alarm
setpoint. Currently Not Used.
Maintenance and Troubleshooting | 4-39