Component
Heatsink
Optical Systems
O-ring
Sample Inlet/Outlet
Pressure Regulator
Purge Air (Gas)
Purge/Shut In
Solenoid Valve
Relief Valve
Bypass Valve
Vent Line
Electrical Leads
1-8 | 241CE II Hydrocarbon Dewpoint Analyzer
Function
Absorbs heat from the Mirror during the Cooling Stage. The heat is then
transferred to the atmosphere through the Heatsink fins. During the Warming
Stage, the Heatsink serves as a passive source of heat by using the surround-
ing ambient temperature to heat the mirror.
The 241CE II Analyzer employs two Optical Systems in its Measuring Cell:
• One detects hydrocarbon condensation on the rough, black surface.
• One detects water condensation on the smooth, reflective surface. (This
water dewpoint measurement is for internal use only, used for compensa-
tion in the calculation of HCDP measurements.)
For each optical system, light from a light-emitting diode (LED) is directed to
a mirror surface through a Polycarbonate Window. The optic-path windows
are truncated cones and thus will remain seated firmly even with high-
pressure gas samples in the Measuring Cell. The placement of the optic-path
windows and Phototransistors are not symmetric, as suggested by Figure 1-3.
These elements are located, by design, so that when condensation occurs
– either hydrocarbon on the rough surface, or water on the smooth surface –
the intensity of light at the respective phototransistor increases. This arrange-
ment is possible because of the different wetting properties of the two types
of liquids.
The gas sample chamber is formed by the Mirror Support and the cavity
machined within the aluminum cylinder. The gas seal between the two parts
is created by an o-ring.
The Sample Gas Inlet and Outlet channels are simply bored through the alu-
minum cylinder. Both the Sample Inlet and Outlet connections contain flame
arrestors.
Used to adjust the sample gas flow rate. This regulator is labelled "Flow Ad-
justment" on the analyzer backpan.
The purge air (gas) supply creates a dry atmosphere within the Thermoelec-
tric Cooler (TEC) compartment to prevent ice from forming, which can dam-
age the cooler.
Controls the flow of the sample gas. At the start of the Purging/Hold Stage,
the solenoid valve is opened to allow the sample gas to purge through the
Measuring Cell. At the end of the Purging/Hold Stage, the solenoid valve is
closed to stop the flow and to enable the Measuring Cell to equilibrate to the
process stream pressure.
The overpressure Relief Valve, which is set to release at 13.8 BARG (200 PSIG),
prevents the solenoid valve from becoming over-pressurized (attaining full
line pressure) in the case of a failure of the pressure regulator.
The Bypass Valve allows a small flow rate during the measuring cycle. The
flow rate through the bypass is approximately one-tenth of the Purge/Hold
Stage cycle flow (called a "trickle purge").
Vents the used sample gas to atmosphere or to an approved low pressure
(<50 PSIG) flare header.
The analyzer is controlled by a single Microcontroller board which, along
with the power supplies and other electronic components, is contained in an
enclosure separate from the Measuring Cell Assembly (Figure 1-4). Communi-
cation between the Microcontroller board and the Measuring Cell Assembly
is via Electrical Leads which pass through the Measuring Cell seal mounting
assembly. The Microcontroller monitors the mirror temperature, controls the
mirror cooling rate, adjusts the brightness level of the LEDs, monitors the
light intensities at the phototransistors, controls the sample valve, and deter-
mines whether a dewpoint is detected.