Checking Flamepaths And Gaps; Checking And Recording Operating Temperatures - Ametek 241CE II User Manual

Hydrocarbon dewpoint analyzer
Hide thumbs Also See for 241CE II:
Table of Contents

Advertisement

RUN Mode
PN 903-8797, Rev A
Before performing maintenance on the analyzer, bleed down the pres-
sure in the sample system by closing the isolation valve on the Sample
Probe or at the Sample Inlet valve. Then, open the Bypass Valve, if not
already done, and allow the gas pressure in the sample system and
Measuring Cell to depressurize for approximately 5 minutes.
This procedure is necessary prior to changing out any components
or replaceable parts, as part of regular preventive or general mainte-
nance.

Checking Flamepaths and Gaps

Possible Cause/
Condition
Corrective Action
Electronics Enclosure
• Flamepath joining
surfaces have scratches,
Each time the Electronics Enclosure is opened, inspect the
dents, or other damage.
joining surfaces for damage (scratches, dents, etc.).
• Flamepaths around
Before closing the Electronics Enclosure door, carefully clean the
User Interface win-
joining surfaces (flamepath) with a soft non-abrasive cloth.
dows/push-buttons are
During each analyzer maintenance, use a feeler gauge to
worn or damaged.
check the flamepath gap of the Electronics Enclosure flanges
(enclosure door and housing joining surfaces). The gap must
not exceed 0.08 mm; if the gap exceeds this value, contact
AMETEK for advice.
See "Examining and Caring For the Flamepaths" in this chapter.
User Interface Window/Push-buttons
Inspect the flamepaths around the User Interface's window and
push-buttons for signs of wear (discoloration) or damage (bent
or scratched push-buttons).
For information on what to do if these flamepaths appear worn
or damaged, contact AMETEK Service.
See "Examining and Caring For the Flamepaths" in this chapter.
Other Flamepaths
For information on the required dimensions of other flamepath
joints, contact AMETEK Service.
See "Examining and Caring For the Flamepaths" in this chapter.

Checking and Recording Operating Temperatures

From the RUN Mode screen, check and record the operating temperatures to
ensure they are all within 2 % of their setpoints. Compare the current tempera-
tures to the actual settings.
Maintenance and Troubleshooting | 4-7