The Stat\History menu items (listed in their viewing order) and their descrip-
tions are:
Sub-Menu Item
Description
En\Dis Alrm?
This prompt asks if you wish to enable\disable the out-of-specification (Off-Spec) alarm.
To enable the Off-Spec Alarm, press '+' . To disable the Off-Spec Alarm, press '–' .
Clear Alarm?
This prompt asks if you wish to clear (reset) all current alarms (Off-Spec or SCode).
To reset all current alarms, press '+' . The message "Cleared" will be displayed. If you press '–' ,
"Not Cleared" will be displayed.
SCode z
Basic Status Code and messages, where 'z' = the basic SCode (see below) along with its
corresponding message (if the alarm condition exists). If the alarm condition does not exist,
"OK" is displayed (e.g., SCode.# OK, where '#' is the number corresponding with the SCode
message).
The basic SCodes and corresponding messages are listed below. The 'w' or 'f' preceding a
message indicates a 'w'arning or 'f'ault alarm condition.
XSCode z
Extended Status Codes and messages, where 'z' = the XSCode (see below) along with its
corresponding message (if the alarm condition exists). If the alarm condition does not exist,
"OK" is displayed (e.g., XSCode.# OK, where '#' is the number corresponding with the XSCode
message).
The XSCodes and corresponding messages are listed below. The 'w' or 'f' preceding a message
indicates a 'w'arning or 'f'ault alarm condition.
3-12 | 241CE II Hydrocarbon Dewpoint Analyzer
Primary
Error
Code
Code #
Message [Description]
w
TimeOut [Stage-Time-Out]
1
1
2
2
w
No H2O DP [Water-Dewpoint-Temperature-Not-Found] –
Available only if H2O DP:INT option is enabled from the
MonitorCFG menu.
w
No HC DP [Hydrocarbon-Dewpoint-Temperature-Not-Found]
4
3
8
4
w
Check XSCode [Advice – Review-Extended-Status-Code]
w
Dirty Cell [Possible-Cell-Contamination]
16
5
32
6
f
H2O Optics [Problem-In-Water-Side-Optical-Path]
f
HC Optics [Problem-In-Hydrocarbon-Side-Optical-Path]
64
7
128
8
f
Cool Cct [Problem-In-Cooling-Circuit]
Primary
Error
Code
Code #
Message [Description]
w
EEP Warn [End-Of-EEPROM-Life]
1
1
2
2
w
Hold TimeOut [Purging/Hold-Stage-Time-Out]
w
Cool TimeOut [Cooling-Stage-Time-Out]
4
3
8
4
w
Warm TimeOut [Warming-Stage-Time-Out]
w
Cross Talk [Excessive-Mirror-Cross-Talk]
16
5
32
6
w
High DP [Hydrocarbon-Dewpoint-Temperature-Too-High]
f
High Temp [Mirror-RTD-Temperature-Is-Too-High]
64
7
f
Low Temp [Mirror RTD-Temperature-Is-Too-Low]
128
8
256
9
f
H2O Dim [Light-Intensity-On-Water-Side-Is-Too-Low]
f
HC Dim [Light-Intensity-On-Hydrocarbon-Side-Is-Too-Low]
512
10
PN 903-8797, Rev A