5. Repeat Steps 4-b and 4-c to ensure the readings on the User Interface
PN 903-8797, Rev A
d. Use the following formulas to calculate a new ExLo and ExHi to com-
pensate for the differences between the input value and the displayed
value:
A = Range Lo (ExLo range [low output from the external water dewpoint/
water content source])
B = Range Hi (ExHi range [high output from the external water dewpoint/
water content source])
C = Reading Lo (external water dewpoint/water content reading on the
User Interface)
D = Reading Hi (external water dewpoint/water content reading on the
User Interface)
E = New ExLo (newly calculated ExLo reading, from Equation 1).
New ExLo =
New ExHi =
e. If the new ExLo and new ExHi values are different, enter these values in
to the analyzer.
match the input signal.
(
C
A
)
x
(
B
A
)
A
(
)
D
C
(
B
A
)
x
(
B
A
)
E
+
(
D
C
)
Appendix A
Equation 1
Equation 2
|
A-5