The User Interface is made up of two areas:
3-2 | 241CE II Hydrocarbon Dewpoint Analyzer
Display Screen/LEDs
Consists of one self-illuminated line capable of displaying up to 12 alpha-
numeric characters and one row of seven LEDs. The information displayed
depends on the current operation mode.
-
The self-illuminated line displays the results of the most recent measure-
ment cycle, mirror temperature, run time, and Status Code (SCode).
To change an operating parameter, it must be active on the display. The
display line flashes when the 241CE II Analyzer is in Standby Mode (a
Fault alarm is present).
-
The LEDs provide a visual indicator of the built-in diagnostics function.
When a LED is On, it indicates the alarm condition associated with the
LED currently exists. For more information about this diagnostics tool,
see "Types of Alarms" in Chapter 4.
The LEDs, numbered 1–7 from left to right, and their conditions when
turned On are:
LED #
Setpoint By
1
User
2
Diagnostics
3
Diagnostics
4
Diagnostics
5
User
6
User
7
User
Condition Existing When LED Turned On
The Hydrocarbon Dewpoint temperature Off-Spec
Alarm is disabled (Stat\History : En\Dis Alrm?).
Fault Alarm: Analyzer in Standby Mode – Immediate
Service Required (review SCode and XSCode in the
Stat\History menu to gather information before
contacting AMETEK). The alphanumeric display will
also flash to alert the operator to this condition.
Warning Alarm: Analyzer reporting an abnormal
condition – Investigation Required (review SCode and
XSCode in the Stat\History menu and attempt to
correct the problems).
Hydrocarbon Dewpoint Temperature Not Found (NoD),
or Purging/Hold Stage Is In Progress (Pur). This setting
is made in the ApplicCFG : UsrRelay menu.
See also the UsrRelay description under "Working in
the Application Configuration Menu" in this chapter.
Reserved for Water Content OffSpec Alarm setpoint.
Currently Not Used.
The current Hydrocarbon Dewpoint temperature is
higher than the Hydrocarbon Dewpoint temperature
Off-Spec Alarm setpoint value (ApplicCFG : HCDPA).
This alarm will only be seen if the Off-Spec Alarm is
enabled in the Stat\History menu (LED 1 is Off when
the Off-Spec Alarm is enabled).
Reserved for Water Dewpoint temperature OffSpec
Alarm setpoint. Currently Not Used.