Advertisement

Sub-Menu Item
Description
Water Measurement
(Continued)
Option
• If H2O DP:INT is enabled, select an option from the Options code sub-menu:
• If H2O DP:DISAB is enabled, select an option from the Options code sub-menu:
Options
The Options code sub-menu allows you to select certain parameters and whether to display
their values in Metric or Imperial measurements. Note that the current selection will be
displayed next to Options on the display (e.g., "Options DP&m"). Options include:
• DP&m = Dewpoint(s), metric
• DP&i = Dewpoint(s), imperial
• WC&m = Water Content, metric
• WC&i = Water Content, imperial
• PMVm = Parts per million by volume, metric
• PMVi = Parts per million by volume, imperial
• PMWm = Parts per million by weight, metric
• PMWi = Parts per million by weight, imperial
Note:
PLo
Pressure input low scale (KPA or PSI). If the pressure signal is supplied from an external source,
the PLo parameter can be used to input the range of the pressure signal corresponding to
4.00 mA. This input is currently not used.
PHi
Pressure input high scale (KPA or PSI). If the pressure signal is supplied from an external
source, the PHi parameter can be used to input the range of the pressure signal
corresponding to 20.00 mA. This input is currently not used.
ExHi
External water content input high scale (°C, °F, mg/m
Note:
ExLo
External water content input low scale (°C, °F, mg/m
Note:
PN 903-8797, Rev A
- DP&m (dewpoint, metric units).
- DP&i (dewpoint, imperial units).
- WC&m (water content, metric units).
- WC&i (water content, imperial units).
- DP&m (dewpoint, metric units).
- DP&i (dewpoint, imperial units).
[if H2O DP:EXT, H2O DP:INT, or H2O DP:DISAB is enabled]
[if H2O DP:EXT, H2O DP:INT, or H2O DP:DISAB is enabled]
[if H2O CT:EXT or H2O DP:INT is enabled]
[if H2O CT:EXT or H2O DP:INT is enabled]
[if H2O CT:EXT is enabled]
[if H2O CT:EXT is enabled].
[if H2O CT:EXT is enabled]
[if H2O CT:EXT is enabled]
You can change the Options code parameter at any time, but first carefully consider
what the displayed results will be. Example, if H2O DP:EXT is selected from the Water
Measurement Option sub-menu, and you have PMVm selected from the Options
code sub-menu, the required results will not be displayed (water dewpoint cannot be
displayed in parts per million by volume).
This signal will only be displayed if an external water signal is provided and the
H2O DP:EXT or H2O CT:EXT option is enabled from the MonitorCFG menu.
This signal will only be displayed if an external water signal is provided and the
H2O DP:EXT or H2O CT:EXT option is enabled from the MonitorCFG menu.
3
, lb/MMcf, PPMV, PPMW).
3
, lb/MMcf, PPMV, PPMW).
User Interface
|
3-19