Ametek 241CE II User Manual page 83

Hydrocarbon dewpoint analyzer
Hide thumbs Also See for 241CE II:
Table of Contents

Advertisement

Sub-Menu Item
Description
I1H
On the first analog output – V/I #1 – this is the full-scale (maximum) WDP temperature
(°C or °F) corresponding to the 20.00 mA output.
Note:
I2L
On the second analog output – V/I #2 – this is the low-scale (minimum) HCDP temperature
(°C or °F) corresponding to the 4.00 mA output.
On the second analog output – V/I #2 – this is the full-scale (maximum) HCDP temperature
I2H
(°C or °F) corresponding to the 20.00 mA output.
On the third analog output – V/I #3 – this is the low-scale (minimum) WCT – mg/m
I3L
PPMV, or PPMW – corresponding to the 4.00 mA output.
Note:
On the third analog output – V/I #3 – this is the full-scale (maximum) WCT – mg/m
I3H
PPMV, or PPMW – corresponding to the 20.00 mA output.
Note:
I4L
The fourth analog output – V/I #4 – is reserved for the low-scale (minimum) pressure
(KPA or PSI) corresponding to the 4.00 mA output. This analog output is currently not used.
I4H
The fourth analog output – V/I #4 – is the full-scale (maximum) pressure (KPA or PSI)
corresponding to the 20.00 mA output. This analog output is currently not used.
This prompt asks if you wish to change Passcode1. To change the password for the ApplicCFG
Chng PSCD1?
menu, enter a new 4-character code, using any combination of the , '+' , and '–' buttons. After
entering the password, "Chng PSCD1?" will be displayed again. Enter the passcode a second
time.
Save AppCFG?
This prompt asks if you wish to permanently save the changes you made to the ApplicCFG
parameters. To save any changes you made, press the '+' button. If the changes are not saved,
they will remain in effect until the analyzer is powered down or reset.
PN 903-8797, Rev A
This parameter will be displayed only if an external water signal is provided and if
the H2O DP:EXT option is enabled from the MonitorCFG menu.
This parameter will be displayed only if an external water signal is provided and if
the H2O CT:EXT option is enabled.
This parameter will be displayed only if an external water signal is provided and if
the H2O CT:EXT option is enabled.
3
(lb/MMcf ),
3
(lb/MMcf ),
|
User Interface
3-17