The status codes in Example 2 indicate:
The combination of status codes in Example 2 indicates:
The necessary corrective actions for Example 2 are:

Suggested Corrective Actions

Most of the alarms, SCodes, and XSCodes are for informational purposes or
are warnings of minor occurrences during the measurement cycle that may
be limiting the capability or range of the analyzer. Although these warnings
should be investigated, all of them do not always require corrective action.
For those that do, suggested corrective actions to correct the problems are
provided in the SCodes and XSCodes troubleshooting charts. A fault alarm
always requires corrective action, even if it is only to reset the alarms. If the
problem persists after following these suggestions, contact AMETEK Service.
4-44 | 241CE II Hydrocarbon Dewpoint Analyzer
The light paths are obscured by debris or other matter.
The hydrocarbon dewpoint temperature is very high.
The Thermoelectric Cooler could not cool the mirror to its Low setpoint
temperature.
The mirror could not warm up to its High setpoint temperature.
The surrounding ambient temperature is too low or the High setpoint
temperature for the mirror is too high, and
The Measuring Cell is contaminated, probably due to being flooded with
liquid hydrocarbons.
Reduce the High setpoint temperature, if necessary, or upgrade the heat-
ing of the enclosure/shelter in which the analyzer is installed.
Clean the Measuring Cell, clean the Filterblock Assembly, replace the Mem-
brane Filters, and clean or replace the Sample Lines.
PN 903-8797, Rev A