Sub-Menu Item
Description
This parameter displays the duration each signal will pause on the RUN menu screen during
Scroll
the auto-scroll (factory default = 4 seconds).
V/I Analog output #1 Zero calibration value, WDP (4.00 mA).
I1Z
Note:
V/I Analog output #1 Span calibration value, WDP (20.00 mA).
I1S
Note:
I2Z
V/I Analog output #2 Zero calibration value, HCDP (4.00 mA).
I2S
V/I Analog output #2 Span calibration value, HCDP (20.00 mA).
I3Z
V/I Analog output #3 Zero calibration value, WCT (4.00 mA).
Note:
I3S
V/I Analog output #3 Span calibration value, WCT (20.00 mA).
Note:
I4Z
V/I Analog output #4 Zero calibration value, Status Code (4.00 mA).
I4S
V/I Analog output #4 Span calibration value, Status Code (20.00 mA).
Note:
Chng PSCD2?
This prompt asks if you wish to change Passcode2. To change the password for the
MonitorCFG menu, enter a new 4-character code, using any combination of the , '+' , and
'–' buttons. After entering the password, "Chng PSCD2?" will be displayed again. Enter the
password a second time.
Save MonCFG?
This prompt asks if you wish to permanently save the changes you made to the MonitorCFG
parameters. To save any changes you made, press the '+' button. If the changes are not saved,
they will remain in effect until the analyzer is powered down or reset.
PN 903-8797, Rev A
This parameter will be displayed only if an external water signal is provided and if
the H2O DP:EXT option is enabled from the MonitorCFG menu.
This parameter will be displayed only if an external water signal is provided and if
the H2O DP:EXT option is enabled from the MonitorCFG menu.
This parameter will be displayed only if an external water signal is provided and if
the H2O CT:EXT option is enabled from the MonitorCFG menu.
This parameter will be displayed only if an external water signal is provided and if
the H2O CT:EXT option is enabled from the MonitorCFG menu.
See "Current Output Zero/Span Calibration," following this section, for
information on how to calibrate the Zero/Span V/I outputs.
|