Patient Data - Respironics V60 User Manual

Ventilator
Table of Contents

Advertisement

Patient data

Table 11-3: Patient data ranges, resolutions, and accuracies during ventilation
Parameter
Breath phase/trigger indicator
PIP
Pt. Leak
Pt. Trig
Rate
T
/T
I
TOT
Tot.Leak
E
V
T
P waveform
waveform
V waveform
Range
Patient data window
Spont, Timed, Exhale
0 to 50 cmH
O
2
0 to 200 L/min BTPS
0 to 100%
0 to 90 BPM
0% to 91%
0 to 200 L/min BTPS
0 to 99.0 L/min BTPS
0 to 3500 mL BTPS
Waveform window
0 to 50 cmH
O
2
-240 to 240 L/min BTPS
0 to 3500 mL BTPS
Resolution
Color-coded display: Spont -
turquoise, Timed - orange, Ex-
hale - blue
1 cmH
O
2
1 L/min
1%
1 BPM
1%
1 L/min
0.1 L/min
1 mL
Time axis: 1 second
Time axis: 1 second
Time axis: 1 second
Technical specifications
Accuracy
N/A
± 2 cmH
O
2
N/A
± 10%
± 1 BPM
± 5%
N/A
± 15% or 0.3 L/min (whichev-
er is greater)
± 15% for volumes above
200 mL
Note: Accuracy specification
was measured with patient
circuit PN 582073
N/A
N/A
N/A
11-3