Technical Specifications; Control Settings - Respironics V60 User Manual

Ventilator
Table of Contents

Advertisement

Chapter 11. Technical specifications

Control settings

Table 11-1: Control setting ranges, resolutions, and accuracies
Parameter
Range
Modes
AVAPS, CPAP, S/T, PCV,
PPV (optional)
C-Flex
OFF, 1 to 3
CPAP
4 to 25 cmH
EPAP
4 to 25 cmH
Flow
10 to 80 L/min
(High flow therapy,
3.00 software and
above, and V60 Plus)
IPAP
4 to 40 cmH
I-Time (Inspiratory
0.30 to 3.00 sec
Time)
Max E
0 to 100 cmH
Max P (PPV Maximum
5 to 40 cmH
Pressure Limit)
Max P (AVAPS Maxi-
6 to 40 cmH
mum IPAP Pressure)
Max R
0 to 50 cmH
Max V (PPV Maximum
200 to 3500 mL
Volume Limit)
Min P (AVAPS Mini-
5 to 30 cmH
mum IPAP Pressure)
O
(Oxygen)
21 to 100%
2
PPV %
0 to 100%
Table 11-1 lists ventilator control setting ranges, resolutions, and accuracies.
Table 11-2 lists the controls active in the different ventilation modes.
Resolution
Mode settings
N/A
Control settings
1
O
1 cmH
O
2
2
O
1 cmH
O
2
2
5 L/min
O
1 cmH
O
2
2
0.05 sec
O/L
1 cmH
O/L
2
2
O
1 cmH
O
2
2
O
1 cmH
O
2
2
O/L/sec
1 cmH
O/L/s
2
2
5 mL
O
1 cmH
O
2
2
1%
1%
Performance Accuracy
Factory default
N/A
S/T
N/A
2
± (2 cmH
O + 4% of
4 cmH
2
target)
± (2 cmH
O + 4% of
4 cmH
2
target)
N/A
35 L/min
± (2 cmH
O + 4% of
12 cmH
2
target)
± 0.03 sec
1.00 sec
N/A
15 cmH
± (2 cmH
O + 4% of
20 cmH
2
target)
± (2 cmH
O + 4% of
25 cmH
2
target)
N/A
4 cmH
± 15%
1000 mL
± (2 cmH
O + 4% of
10 cmH
2
target)
± 5%
21%
N/A
30%
O
2
O
2
O
2
O/L
2
O
2
O
2
O/L/s
2
O
2
11-1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

V60 plus

Table of Contents