2. Select Control → Auto Sequence... to start auto sequencing of the line gas
Rosemount 700XA
Unless stated otherwise in the product documentation, ensure that the
pressure of the calibration and sample line is regulated at 10 to 30 psig (0.7
to 2.1 BarG). 15 psig (1 BarG) is recommended.
c) Validate calibration gas and retention times and run a manual calibration.
d) Go to MON2020 → Application → Component Data and select the
associated stream. Check the Component Data table for calibration gas
validation information and retention times.
e) Go to MON2020 → Control → Calibration and select the analysis stream to
run a manual calibration. Select the Purge stream for 60 seconds checkbox
and Normal calibration type radio button; then click OK.
NOTICE
Consult the
Rosemount MON2020 Software for Gas Chromatographs
Reference Manual
stream(s).
NOTICE
Consult the
Rosemount MON2020 Software for Gas Chromatographs Reference
Manual
for more information.
The gas chromatograph (GC) begins the auto sequence analysis.
for more information.
Installation and start-up
June 2022
99