Emerson Rosemount 700XA Reference Manual page 175

Gas chromatograph
Hide thumbs Also See for Rosemount 700XA:
Table of Contents

Advertisement

Reference Manual
2-3-9000-744
A F
A F
Two screws
Two flat washers
Remove a F
Note
Be sure to properly ground yourself before performing this procedure.
Procedure
1. Unscrew the two local operator interface (LOI) post tips.
2. Remove the two screws that attach the F
Install a F
NOTICE
Properly ground yourself before performing this procedure.
Note
The gas chromatograph (GC) draws 21 mA from the F
Procedure
1. Attach the Fieldbus assembly bracket to the F
2. Attach the Fieldbus assembly bracket to the card cage by aligning the Fieldbus
3. Screw in the LOI post tips.
4. Use the
Rosemount 700XA
Fieldbus assembly bracket
OUNDATION
Fieldbus cable assembly
OUNDATION
OUNDATION
You can now detach the F
the F
Fieldbus module.
OUNDATION
OUNDATION
aligning the two holes in the Fieldbus assembly bracket with the two holes at the
bottom of the F
OUNDATION
assembly bracket's second set of holes with the local operator interface (LOI) post
tip holes in the card cage.
Table 5-4
to connect the F
backplane:
Table 5-4: F
Fieldbus wiring
OUNDATION
Backplane terminal block
TB15
TB13
Fieldbus module
Fieldbus module from the card cage.
OUNDATION
OUNDATION
Fieldbus module
OUNDATION
OUNDATION
Fieldbus module and screwing in the two thumb screws.
Fieldbus cable assembly to the
OUNDATION
Post number
1
2
3
3
4
Operation and maintenance
June 2022
Fieldbus assembly bracket to
Fieldbus module.
Fieldbus module by
Wire color
Brown
White
Green
Red
Black
175