Installation and start-up
June 2022
6. Access the backplane, which is located in the GC's lower enclosure.
7. If you are setting up a static IP address for the Ethernet1 port at J22, and you also
8. To connect to the GC:
74
Figure 4-18: Port locations on the backplane
intend to connect to your company's local area network, do the following:
a) Locate the set of dip switches, labeled 1 and 2 , at SW1 on the backplane.
SW1 is located directly beneath the Ethernet port at J22 .
b) Move dip switch 1 to its left position (Off).
This disables the DHCP server.
a) Start Rosemount MON2020 and select File → GC Directory....
The GC Directory window displays.
b) Select Add.
Rosemount MON2020 adds a new GC profile to the end of the table.
Note
You can name the GC's profile as well as add a short description.
c) Select the new profile and click Ethernet... Enter the GC's static IP address in
the IP address field.
d) Click OK.
The Ethernet Connection Properties for New GC window closes.
Reference Manual
2-3-9000-744
Emerson.com/Rosemount