Local operator interface (LOI)
June 2022
6. Click F1.
7. Click F1 (SELECT) and navigate up or down the list to highlight your user name.
8. Click ENTER.
9. Navigate to the Password field, press F1 (EDIT), and enter your password.
214
You must be logged in to the gas chromatograph (GC) to make a change to any
screen. If you try to edit a field before logging in—as you just did, the LOI displays
the Login dialog to prompt you to log in.
Figure A-12: Login screen (Edit mode prompt)
Note
Notice that there is also a navigation icon on the Login dialog.
Note
For the remainder of this tutorial, when referring to the F1 key, the key's current
valid function will be indicated in parenthesis—for instance, F1 (MOVE) or F1
(SELECT).
Important
The password security policy requires a login name and password. Security levels
include:
High - password policy level includes following parameters:
Ensure minimum eight characters length and maximum of 12 characters.
Ensure at least one upper case character.
Ensure at least one lower case character.
Ensure at least one digit.
Ensure at least one special character: