Reference Manual
2-3-9000-744
If there is reason to suspect that the span on any particular channel might be off after a
period of time and heavy use, then recalibrate the analog output for that channel.
Analog output adjustment
Emerson sets the initial analog output adjustments at the factory before shipment at
standard values (4-20 mA). You many need to check and/or adjust these values depending
on output cabling/impedance.
The adjustment may require two people if the units are some distance apart. It requires a
calibrated digital meter to check the zero and full scale values at the receiving end. You
can use MON2020 to adjust the scale or span value.
You can use different engineering units, volts, and percentages to calibrate the analog
outputs.
5.4.15
Upgrading the embedded software
The base GC firmware performs functions similar to operating systems such as DOS,
Windows
run the customer's applications. There is no direct user-level interface to the firmware.
If you need to upgrade the firmware to your system, refer to the
Gas Chromatographs Reference Manual
The GC's applications use the tools provided by firmware to perform the desired gas
chromatograph applications for the user. There are different applications to facilitate
different gas chromatographic needs. To load a new application or to upgrade an existing
application, refer to the
for details.
Rosemount 700XA
®
, or Linux
. The base GC firmware provides the basic resources and interfaces to
MON2020 Software for Gas Chromatographs Reference Manual
Operation and maintenance
MON2020 Software for
for additional information.
June 2022
181