Installation and start-up
June 2022
Terminal
1
2
3
4
5
6
7
8
9
10
To connect the ROC800 DO module to a field device:
Procedure
1. Expose the end of the wire to a maximum length of 0.25 in (6.4 mm).
2. Insert the exposed end into the clamp beneath the termination screw.
3. Tighten the screw.
82
Label
1+
COM
2+
COM
3+
COM
4+
COM
5+
COM
NOTICE
We recommend twisted-pair cables for in/out (IO) signal wiring. The module's
terminal blocks accept wire sizes between 12 and 22 American wire gauge (AWG).
Allow some slack when making connections to prevent strain.
NOTICE
Failure to follow this notice may cause a short circuit and damage equipment. Allow
only a minimal amount of bare wire to prevent short circuits.
Definition
Positive discrete output
Discrete output return
Positive discrete output
Discrete output return
Positive discrete output
Discrete output return
Positive discrete output
Discrete output return
Positive discrete output
Discrete output return
Reference Manual
2-3-9000-744
Emerson.com/Rosemount