Emerson Rosemount 700XA Reference Manual page 148

Gas chromatograph
Hide thumbs Also See for Rosemount 700XA:
Table of Contents

Advertisement

Operation and maintenance
June 2022
8. Disconnect the wire terminals for J6, J10, and J2.
9. Loosen the two Phillips screws from the stream solenoid manifold.
10. Move the entire manifold away from the µFPD PMT module.
11. Use a large flat head screwdriver to loosen the two µFPD PMT module bracket
12. Move the 1/16-in. stainless steel tubing away and pull the µFPD PMT module out of
Install the micro flame photometric detector (µFPD) photo
multiplier tube (PMT) module
Tools required
7 ⁄ 64-in. Allen wrench
Phillips screwdriver
Flathead screwdriver
Procedure
1. Lower the µFPD PMT module into the enclosure.
148
Figure 5-34: µFPD PMT module connections
A. J6 - CAN and power signals
B. J10 - igniter
C. J2 - thermocouple
D. USB connector
E. PMT module bracket screws
F. Stream manifold screws
G. Stream manifold
screws.
the enclosure.
Reference Manual
2-3-9000-744
Emerson.com/Rosemount

Advertisement

Table of Contents
loading

Table of Contents