Figure 5-40: Methanator assembly
A. Methanator housing cover
B. Heater insulator (2)
C. Heater insulator, split
D. Heater block, methanator bottom
E. Methanator column
F. RTD
G. Cartridge heater
H. Heater block, methanator top
I. Methanator housing
The RTD is replaceable. When replacing it, take care to anchor the RTD cable to the tubing
to prevent loosening over time.
To replace the RTD, consult drawing #CE-22715.