Configure The Micro Flame Photometric Detector (Μfpd) - Emerson Rosemount 700XA Reference Manual

Gas chromatograph
Hide thumbs Also See for Rosemount 700XA:
Table of Contents

Advertisement

Micro flame photometric detector (µFPD)
June 2022
Figure C-2: µFPD PMT
C.1
Configure the micro flame photometric
detector (µFPD)
Procedure
1. Open MON2020 and select Hardware → Detectors.
2. Select Hardware → Detectors.
3. Select Manual if you want to control the burner ignition; select Auto if you want the
Name
Det #
Detector
H2 Valve
Flame Temp RTD
268
If your gas chromatograph (GC) has an integral FPD, FPD G2 displays in Slot 1 or Slot
2.
Note
Before making any modifications to this window, halt the analysis.
GC to control the burner ignition.
The following data displays for each detector:
Description
Numerical identifier for the detector to which the following data applies.
Options, which depend on your GC's configuration, are:
TCD (thermal conductivity detector)
FPD G2 (µFPD)
FID (flame ionization detector)
FID G2
Optional hydrogen carrier shut-off valve
Select the appropriate resistance temperature device (RTD) from the drop-down
list. The RTD measures the temperature of the flame.
Reference Manual
2-3-9000-744
Emerson.com/Rosemount

Advertisement

Table of Contents
loading

Table of Contents