Emerson Rosemount 700XA Reference Manual page 201

Gas chromatograph
Hide thumbs Also See for Rosemount 700XA:
Table of Contents

Advertisement

Reference Manual
2-3-9000-744
Screen
Heater
Valves
In Status Display mode, pressing the RIGHT key takes you to the next available non-idle
analysis live CGM screen. If all the cycle clocks are idle, then press RIGHT on the Status
screen to go to the Alarm screen. Press LEFT on the Alarm screen to go to the Status screen.
When you press RIGHT, the LOI displays each screen in succession and then loops back to
the Status screen.
Figure A-2: LOI Status screen
This screen displays the Analysis names and all the four cycle clocks in the Status table. The
LOI status variables are shown for one cycle clock every 15 seconds as depicted in
A-2. For the next 15 seconds, the LOI automatically scrolls and displays the next cycle
clock's variables.
In Status Display mode, you can manually scroll to the next screen using the RIGHT key, or
to the previous screen using the LEFT key. You can pause automatic scrolling at any time
by pressing the EXIT key, and you can resume automatic scrolling by pressing either the
LEFT or the RIGHT key. Automatic scrolling resumes after ten minutes of keypad inactivity.
Press F1 when MOVE is displayed in the green box below to foucs inside the screen, so that
you can navigate through the controls of the screen using the LEFT, RIGHT, UP, and
DOWN keys. Press EXIT to return focus to the top level (outside of the screen). Press LEFT
Rosemount 700XA
Description
Displays information about the PID temperature control loops.
Displays the settings and states of the stream and analyzer valves.
Local operator interface (LOI)
June 2022
Figure
201