Operation and maintenance
June 2022
5.4.12
Communications
The Rosemount 700XA has four serial communication ports: Port 0, Port 1, Port 2, and
Port 3, which is a dedicated personal computer (PC) to gas chromatograph (GC) port.
You can set the mode for each of the first three ports to RS232, RS422, or RS485.
Normally, the customer specifies these port configurations at the time of order, and
Emerson sets them at the factory. However, the customer can use Rosemount MON2020
to change them at any time.
Note
The backplane has two switches located at SW1. The first switch is used for starting the
DHCP server. The second switch is reserved for future use.
The backplane has two Ethernet ports:
Name
ETHERNET1
ETHERNET2
Related information
Connect directly to a personal computer (PC) using the gas chromatograph's (GC's)
Ethernet1 port
Maximum distance by communication type
Communication Type
RS-232
RS-422/RS-485
Ethernet (CAT5)
164
Location
Connector type
J22
RJ45 (DHCP-enabled)
TB11
4-wire terminal block
Maximum distance
50 ft (15.24 m)
4,000 ft (1,219.20 m)
300 ft (91.44 m)
Reference Manual
2-3-9000-744
Emerson.com/Rosemount