8. Connect the igniter, air intake, hydrogen intake, and sample lines.
9. Replace the thermal hood and the explosion-proof dome.
10. Apply power to the GC.
Rosemount 700XA
WARNING
Burns
Some parts of the analyzer may be heated to 248 °F (120 °C).
To prevent burns, do not touch any of the hot parts. All parts of an analyzer are
always hot unless it has been switched off and allowed to cool down.
Before fitting, removing, or performing any maintenance on the analyzer, make
sure that it has been switched off and allowed to cool for at least two hours.
When handling the analyzer, always use suitable protective gloves.
These precautions are particularly important when working at heights.
If burned, seek medical treatment immediately.
Figure 5-28: µFPD igniter and intake lines
A. Igniter/thermocouple
B. Air intake port
C. Hydrogen intake port
D. Sample intake port
Operation and maintenance
June 2022
141