WARNING
Toxic vapors
Exit ports may discharge dangerous levels of toxic vapors.
Use proper protection and a suitable exhaust device.
WARNING
Burns
Some parts of the analyzer may be heated to 248 °F (120 °C).
To prevent burns, do not touch any of the hot parts. All parts of an analyzer are always hot unless it has been switched off and
allowed to cool down.
Before fitting, removing, or performing any maintenance on the analyzer, make sure that it has been switched off and allowed
to cool for at least two hours.
When handling the analyzer, always use suitable protective gloves.
These precautions are particularly important when working at heights.
If burned, seek medical treatment immediately.
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users' equipment. This could
be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your system. Restrict physical access
by unauthorized personnel to protect end users' assets. This is true for all systems used within the facility.
NOTICE
Replaceable parts
Only a few parts inside the device are replaceable. Only trained service personnel should replace parts.
All replacement parts must be authorized by Emerson to ensure product certification compliance.
NOTICE
Equipment damage
If the device is heated without carrier flow, damage to the columns may occur.
NOTICE
Waste disposal
Waste electrical and electronic products must not be disposed of with household waste.
Please recycle where facilities exist.
Check with your local authority or retailer for recycling advice.
NOTICE
The device is certified by CSA and ATEX. See the certification tag on the device for specific details about its agency approvals.
When the vapor regulators and flow switches are fitted, they must be suitably certified with the ratings Ex d IIC Gb T6/T4/T3 and
for a minimum ambient temperature range: Ta = -20 °C to +60 °C.
Where right angle bend cable adapters are used, they shall be appropriately certified and shall interface with enclosures via
appropriate certified barrier glands.