Installation and start-up
June 2022
2. Start Rosemount MON2020 and do the following to create a GC connection for the
3. Connect one end of the direct connect cable to the GC's serial port at J23 on the
70
The Browse dialog window displays.
f) Navigate to the Rosemount MON2020 install directory (typically C:\Program
Files (x86)\Emerson Process Management\MON2020) and select Daniel Direct
Connection.inf.
g) Click Open.
You return to the Install from Disk dialog window.
h) Click OK.
You return to the Add Hardware Wizard.
i) Click Next.
j) Select an available serial port and click Next.
The Hardware Installation dialog window displays.
k) Click Continue Anyway.
After the driver is installed, you return to the Add Hardware Wizard.
l) Click Finish.
You return to the Phones and Modems dialog window. The Daniel Direct
Connect modem should be listed in the Modem column.
Daniel Direct Connection modem:
a) Go to File → GC Directory....
The GC Directory window displays.
b) Select Add from the GC Directory window's File menu.
A New GC row is added to the bottom of the table.
c) Select the New GC text and type a new name for the GC connection.
d) Select the new GC's Direct check box.
e) Click the Direct button located at the bottom of the GC Directory window.
The Direct Connection Properties window displays.
f) Select Communications cable between two computers (COM n) from the
Port dropdown menu.
Note
The letter n stands for the COM port number.
g) Select 57600 from the Baud Rate dropdown menu.
h) Click OK to save the settings.
You return to the GC Directory window.
i) Click OK to save the new GC connection and to close the GC Directory
window.
backplane.
Reference Manual
2-3-9000-744
Emerson.com/Rosemount