Emerson Rosemount 700XA Reference Manual page 73

Gas chromatograph
Hide thumbs Also See for Rosemount 700XA:
Table of Contents

Advertisement

Reference Manual
2-3-9000-744
4.4.12
Assign a static Internet protocol (IP) address to the gas
chromatograph (GC)
Procedure
1. Start Rosemount MON2020 and log in to the GC using a direct Ethernet connection.
2. Go to Application → Ethernet ports....
3. Depending upon the Ethernet port to which you want to assign a static IP address,
4. Click OK.
5. Log off the GC.
Rosemount 700XA
The Ethernet Ports window displays.
do the following:
a) The Ethernet port at TB11: Enter the appropriate values in the Ethernet2 IP
Address, the Ethernet 2 Subnet, and the Default Gateway fields.
b) The RJ45 Ethernet port at J22: Enter the appropriate values in the Ethernet1
IP Address, the Ethernet1 Subnet, and the Default Gateway fields.
Note
See your information technology (IT) staff to obtain IP, subnet, and gateway
addresses.
Important
To configure a Ethernet IP address using the local operator interface (LOI),
refer to
Figure
A-55.
Installation and start-up
June 2022
73