Reference Manual
2-3-9000-744
6. On the new CPU board, set up switches as shown in the following image:
7. Install the new CPU board in the card cage.
8. Place the clear plastic cover back over the boards.
9. Close the GC cover.
Rosemount 700XA
Figure 5-4: CPU switch settings
A. Turn SW7 ON (toward the dot).
B. Turn SW3 OFF (away from the dot).
C. Turn SW4 OFF (away from the dot).
D. Turn SW6 OFF (away from the dot).
Note
Rosemount 700XA GCs are tagged with CPU board part number 7A00555G02.
Ensure the board is seated firmly in place.