Configure analytical train
Use the Analytical Train Configuration window for multiple analysis clocks to assign the
valves, discrete outputs (DO), detectors, electronic pressure controls (EPCs), and heaters
to each train and then assign each train to its respective analysis clock.
Figure 4-34: Analytical Train Configuration window
Procedure
1. Assign the usage of valves and DO to Analyzer# on Hardware → Valves, Hardware
2. Open the Application → Analytical Train Configuration screen.
3. Click Discrete Output and Valves. Assign the respective DOs, valves, and detectors
Rosemount 700XA
→ Detectors, Hardware → Discrete Outputs, Hardware → EPC, and Hardware →
Heaters screens.
You can use the Filter Selections drop-down list to filter by the type of hardware you
are interested in. The options are:
•
All
•
Detectors
•
Valves
•
Discrete Output
•
EPC
•
Heaters
By default, All is selected, and all types of hardware are displayed. To filter by a
certain type of hardware, select it from the list. Then only rows with the selected
hardware will be displayed.
to each analytical train.
The valves are assigned with Usage as Analyzer# displayed on this screen. All
available detectors are also displayed on this screen. You cannot configure the same
Installation and start-up