Required Tools And Components - Emerson Rosemount 700XA Reference Manual

Gas chromatograph
Hide thumbs Also See for Rosemount 700XA:
Table of Contents

Advertisement

Reference Manual
2-3-9000-744
3.3

Required tools and components

You will need the following tools and components to install the gas chromatograph (GC).
Zero grade carrier gas:
High pressure dual-stage regulator for the carrier gas cylinder
Calibration standard gas with correct number of components and concentrations
Dual-stage regulator for the calibration gas cylinder with a low pressure side capable of
controlling pressure up to 30 psig (2.07 barg)
Sample probe regulator (fixture for procuring the stream or sample gas for
chromatographic analysis)
Coalescing filter
Membrane filter
⅛-in stainless steel tubing
Heat tracing, as required for sample transport and calibration lines
Miscellaneous tube fittings, tubing benders, and tubing cutter
14 American wire gauge (AWG) (18 metric wire gauge [MWG]) or larger electrical
wiring and conduit to provide 120 or 240 Vac, single phase, 50 to 60 Hz, from an
appropriate circuit breaker and power disconnect switch.
Digital volt-ohm meter with probe-type leads
Flow measuring device
Open-end wrenches sized ¼-in, 5/16-in, 7/16-in, ½-in, 9/16-in, and ⅝-in.
Torque wrench
Related information
Gas chromatograph wiring
Rosemount 700XA
99.995% pure
Less than 5 ppm water
Less than 0.5 ppm hydrocarbons
High side up to 3,000 psig (207 barg)
Gauge (psig)
Low side capable of controlling pressure up to 150 psig (10 barg)
For connecting calibration gas to the GC
For connecting carrier gas to the GC
For connecting stream gas to the GC
Sulfinert tubing required if sulfur components are present in calibration gas