7.1.1
Emergency lid release
WARNING
Pos : 167 /010 Univ ers almodul e/
Seitenwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1
50 / 82
In the event of a power failure, it is possible to manually open the centrifuge
lid.
•
Switch off the mains power switch and disconnect the power cord from
the socket.
•
Remove the plug (see figure, item 1) from the opening at the side panel,
e.g. with a screwdriver.
Fig. 28: Position of the opening for the emergency lid release
•
The plug is connected with a string. Pull the string in order to release
the lid.
•
Then, reinsert the stopper into the opening.
Do not unlock or open the lid unless the rotor is at a standstill.
If the lid is opened via the emergency lid release system during a centrifuge
run, the centrifuge will be switched off immediately and decelerate in an
unbraked manner.