Emergency Lid Release - Sigma 3-18KS Operating Manual

Refrigerated centrifuge
Hide thumbs Also See for 3-18KS:
Table of Contents

Advertisement

7.1.1

Emergency lid release

WARNING
Version 08/2012, Rev. 1.26 of 24/10/2023 • sb
Original operating manual, part no. 0701102
In the event of a power failure, it is possible to manually open the centrifuge
lid.
Switch off the mains power switch and disconnect the power cord from
the socket.
Remove the plug (see figure, item 1) from the opening on the right side
panel, e.g. with a screwdriver.
Fig. 33: Position of the opening for the emergency lid release
First, pull the red string to release the mechanism and then unlock the
lid lock by pulling the white string.
Then, reinsert the plug and quit the error message by switching the
centrifuge off/on. Approximately 10 seconds after quitting the lid can be
closed again.
Do not unlock or open the lid unless the rotor is at a standstill.
If the lid is opened via the emergency lid release system during a centrifuge
run, the centrifuge will be switched off immediately and decelerate in an
unbraked manner.
Refrigerated Centrifuge Sigma 3-18KS
7 Malfunctions and error correction
61 / 95