Refrigerated Centrifuge Sigma 2-16KL
8 Maintenance and service
→
Pos : 188 /100 Sigma/100 BA Z entrifugen Sigma (Standardm odul e)/080 War tung und Instandhaltung/080-0010- 0040 Tragbolz en (nur 2- 6, 2-6E, 2- 7, 2-16P, 2-16KL) @ 44\m od_1467889556212_68.docx @ 436524 @ 3 @ 1
8.1.5
Load-bearing bolts
NOTE
Pos : 189 /010 Univ ers almodul e/
Seitenwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1
56 / 82
Plastic bucket
The load-bearing bolts of the rotor must not be greased if plastic buckets
are used!
Aluminium buckets
Only greased load-bearing bolts ensure a uniform swing-out of the buckets
and, therefore, the smooth operation of the centrifuge. Load-bearing bolts
that are insufficiently greased may cause the centrifuge to stop due to an
imbalance.
•
Clean the load-bearing bolts and bucket groove in order to remove the
old grease.
•
Apply a small amount of heavy-duty grease for load-bearing bolts (ref.
no. 71401, see the following picture) to both load-bearing bolts of a
bucket.
Fig. 29: Sufficient quantity of grease for one bolt
•
Install the bucket and swing it manually back and forth once in order to
distribute the grease.
•
Repeat this process with all the other buckets.
Translation of the original operating manual, part no. 0700702
Version 08/2012, Rev. 1.24 of 28/04/2022 • sb