Installation of an angle rotor with a hermetic lid

NOTE
NOTE
Version 08/2012, Rev. 1.24 of 28/04/2022 • sb
Translation of the original operating manual, part no. 0700702
When using rotors for microtiter plate formats:
Ensure that the plate holders are inserted together with the plates into the
buckets.
The lid screw serves for the fastening of the lid onto the rotor only, not for
the fastening of the rotor onto the motor shaft.
Removal of a rotor
Loosen the rotor tie-down screw by turning it anti-clockwise and remove
the rotor.
Open the centrifuge lid by pressing the lid key.
Unscrew the rotor tie-down screw from the motor shaft (counter-
clockwise).
Screw the rotor lid onto the rotor and tighten it by turning it clockwise.
Lower the rotor with the lid with its central bore straight down onto the
motor shaft.
Insert the rotor tie-down screw into the motor shaft. Tighten the rotor tie-
down screw clockwise with the supplied rotor wrench with 3 Nm. In
doing so, hold the rotor at its outer rim.
Follow the safety instructions and hazard warnings (see chapter 3 -
"Safety")!
The rotor can also be used without a cover.
The lid screw serves for the fastening of the lid onto the rotor only, not for
the fastening of the rotor onto the motor shaft.