Installation Of The Microhaematocrit Rotor - Sigma 2-16KL Operating Manual

Refrigerated centrifuge
Table of Contents

Advertisement

Refrigerated Centrifuge Sigma 2-16KL
6 Using the centrifuge
1
Rotor tie down screw
2
Lid
3
Rotor
Pos : 112 /010 Univ ers almodul e/
Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 113 /100 Sigma/110 BA Z entrifugen Sigma (Proj ekte)/2-16KL_2- 16KH L/060 Betrieb/060-0020-0020- 0021 Eins etz en eines Mikr ohämatokrittell ers 2-16KL_KH L @ 29\m od_1405321437508_68.docx @ 196763 @ 4 @ 1
6.2.2.3

Installation of the microhaematocrit rotor

CAUTION
32 / 82
Removal of a rotor
Unscrew the rotor tie-down screw (counter-clockwise) and remove the
rotor with the lid.
Open the lid by turning it counter-clockwise and remove the lid.
Fig. 5: Angle rotor with a hermetic lid
Open the centrifuge lid by pressing the lid key.
Replace the rotor tie-down screw with the rotor tie-down screw of the
microhaematocrit rotor (part no. 81543, included in the scope of supply
of the microhaematocrit rotor), and tighten it loosely.
Lower the rotor with its central bore straight down onto the motor shaft.
Tighten the rotor tie-down screw clockwise with 3 Nm using the supplied
rotor wrench. In doing so, hold the microhaematocrit rotor with one hand
and tilt it slightly in order to prevent the motor shaft from slipping
through.
Check the rotor for a correct and tight fit.
Follow the safety instructions and hazard warnings (see chapter 3 -
"Safety")!
Please consider the maximum permissible gravitational field of 12,000 x g.
An excess gravitational field will result in an increased risk of glass
breakage!
Translation of the original operating manual, part no. 0700702
Version 08/2012, Rev. 1.24 of 28/04/2022 • sb