Customer-provided fuses

NOTE
Pos : 99 /010 U nivers almodule/
Absc hnittsw echs el @ 0\m od_1202124514062_0.docx @ 418 @ @ 1
Pos : 100 /010 Univ ers almodul e/
Seitenwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1
Version 08/2012, Rev. 1.24 of 28/04/2022 • sb
Translation of the original operating manual, part no. 0700702
The centrifuges are equipped with a mains power switch with an integrated
thermal circuit breaker.
Switch the unit off by actuating the mains power switch.
If it has tripped, let the circuit breaker cool for approximately 2 minutes.
Switch the unit on.
The centrifuge is now ready for operation.
Typically, the centrifuge must be protected with 16 Amp B fuses that are to
be provided by the customer.
To ensure safe disconnection in the event of a fault, an AC/DC-sensitive
RCD (residual current device) must be integrated in the wiring system of
the building.
Refrigerated Centrifuge Sigma 2-16KL