7 Malfunctions and error correction
7.1.1

Emergency lid release

WARNING
66 / 125
In the event of a power failure, it is possible to manually open the centrifuge
lid.
Switch off the mains power switch and disconnect the power cord from
the socket.
Remove the plugs (see figure, item 1) at the right side panel, e.g. with a
screw driver.
Fig. 39: Position of the openings for the emergency lid release
Insert the supplied tube wrench (part no. 930 110) horizontally into the
hole. The key will be guided through a funnel-shaped tube to the shaft
of the lid lock motor.
Fig. 40: The emergency lid release key must be inserted horizontally.
Unlock the motorised lid locks as follows:
– Turn the left lid lock anti-clockwise.
– Turn the right lid lock clockwise.
Then, reinsert the plugs.
The lid may only be unlocked and opened when the rotor is at a standstill.
If the lid is opened via the emergency lid release system during a centrifuge
run, the centrifuge will be switched off immediately and decelerate in an
unbraked manner.
Translation of the original operating manual, part no. 0703502