Speed, Radius, And Relative Centrifugal Force; Density - Sigma 2-16KL Operating Manual

Refrigerated centrifuge
Table of Contents

Advertisement

Refrigerated Centrifuge Sigma 2-16KL
2 Layout and mode of operation
Pos : 27 /100 Sigma/100 BA Zentrifugen Sigma (Standar dmodule)/020 Aufbau und Wirk ungsw eise/020-0020-0020- 0010 Dr ehzahl, R adius , R elativ e Zentrifugalbesc hl eunigung @ 26\mod_1405318988082_68.docx @ 191849 @ 4 @ 1
2.2.2.1

Speed, radius, and relative centrifugal force

Pos : 28 /010 U nivers almodule/
Leerz eile @ 0\m od_1202116244500_0.doc x @ 114 @ @ 1
Pos : 29 /100 Sigma/100 BA Zentrifugen Sigma (Standar dmodule)/020 Aufbau und Wirk ungsw eise/020-0020-0020- 0020 Dic hte @ 26\mod_1405318989359_68.docx @ 191863 @ 4 @ 1
2.2.2.2

Density

Pos : 30 /010 U nivers almodule/
Absc hnittsw echs el @ 0\m od_1202124514062_0.docx @ 418 @ @ 1
Pos : 31 /010 U nivers almodule/
Seitenw ec hsel @ 0\m od_1202116244312_0.docx @ 105 @ @ 1
14 / 82
The acceleration g, which the samples are subject to, can be increased by
increasing the radius in the rotor chamber and by increasing the speed.
These three parameters are interdependent and linked with each other via
the following formula:
Relative centrifugal force RCF = 11.18 x 10-6 x r x n
r = radius in cm
n = speed in rpm
RCF without any dimension
If two values are entered, the third value is determined by way of the stated
formula. If, afterwards, the speed or the radius is changed, the resulting
relative centrifugal force will be recalculated automatically by the control
unit. If the RCF is changed, the speed will be adapted while the specified
radius is maintained.
The speed-gravitational-field-diagram provides an overview of the
relationship between speed, radius, and RCF (see chapter 11.2 - "Speed-
gravitational-field-diagram").
The laboratory centrifuge is suitable for the separation of constituents of
different densities in mixtures with a maximum density of 1.2 g/cm
information concerning the speed of rotors and accessories refers to liquids
with a density corresponding to this specification. If the density is above this
value, the maximum permissible speed of the centrifuge must be reduced
based on the following formula:
x √ (1.2/��)
n = n
max
��
3
= density in g/cm
Translation of the original operating manual, part no. 0700702