Version 08/2012, Rev. 1.24 of 28/04/2022 • sb
Translation of the original operating manual, part no. 0700702
Manufacturer
Type:
Connection requirements
Electr. connection:
Protection class:
IP code:
Power consumption (kW):
Input fuse (AT):
Performance data
Max. speed (rpm):
Max. capacity (ml):
Max. gravitational field (x g):
Max. kinetic energy (Nm):
Other parameters
Time range:
Temperature range:
Storage locations:
Physical data
Height (mm):
Height with open lid (mm):
Width (mm):
Depth (mm):
Weight (kg):
Noise level (dB(A)):
Refrigerant data
Refrigerant:
Global warming potential (GWP):
Filling quantity (kg):
Max. permissible pressure (bar):
CO
equivalent (t):
2
Refrigerated Centrifuge Sigma 2-16KL
10 Technical data
Sigma Laborzentrifugen GmbH
An der Unteren Söse 50
37520 Osterode (Germany)
2-16KL
see name plate
I
20
0.72
6.0 (at 220-240 V / 50/60 Hz)
temperature fuses
15,300
400
21,913
9,962
10 sec – 99 h 59 min,
short run, continuous run
-10°C to +40°C
50
310
705
550
570
60
65 (at maximum speed)
R134a
1,300
0.255
23
0.332
63 / 82