Maintenance And Service; Maintenance; Centrifuge - Sigma 2-16KL Operating Manual

Refrigerated centrifuge
Table of Contents

Advertisement

Pos : 173 /100 Sigma/100 BA Z entrifugen Sigma (Standardm odul e)/080 War tung und Inst andhaltung/080 Wartung und Instandhaltung========================= @ 28\m od_1405320612731_68.docx @ 194789 @ 1 @ 1
8

Maintenance and service

CAUTION
Pos : 174 /010 Univ ers almodul e/
Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 175 /100 Sigma/100 BA Z entrifugen Sigma (Standardm odul e)/080 War tung und Instandhaltung/080-0010 W artungsar beiten--------------------------------------- @ 28\mod_1405320613978_68.docx @ 194803 @ 2 @ 1
8.1

Maintenance

Pos : 176 /100 Sigma/100 BA Z entrifugen Sigma (Standardm odul e)/080 War tung und Instandhaltung/080-0010- 0010 Zentrifuge @ 28\mod_1405320615295_68.doc x @ 194817 @ 3 @ 1
8.1.1

Centrifuge

WARNING
Pos : 177 /010 Univ ers almodul e/
Seitenwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1
Version 08/2012, Rev. 1.24 of 28/04/2022 • sb
Translation of the original operating manual, part no. 0700702
The centrifuge, rotor, and accessories are subject to high mechanical
stress. Thorough maintenance performed by the user extends the service
life and prevents premature failure.
If corrosion or other damage occurs due to improper care, the
manufacturer cannot be held liable or subject to any warranty claims.
Use soap water or other water-soluble, mild cleaning agents with a pH
value between 6 and 8 for cleaning the centrifuge and accessories (see
also chapter 8.2 - "Sterilisation and disinfection of the rotor chamber
and accessories").
Avoid corrosive and aggressive substances.
Do not use solvents.
Do not use agents with abrasive particles.
Do not expose the centrifuge and rotors to intensive UV radiation or
thermal stress (e.g. by heat generators).
Unplug the mains power plug before cleaning.
Carefully remove all liquids, including water and particularly all the
solvents, acids, and alkaline solutions from the rotor chamber using a
cloth in order to avoid damage to the motor bearings.
If the centrifuge has been contaminated with toxic, radioactive, or
pathogenic substances, clean the rotor chamber immediately with a
suitable decontamination agent (depending on the type of
contamination).
Take suitable precautions for your own safety if there is a risk of toxic,
radioactive, or pathogenic contamination.
After every cleaning process, grease the motor shaft slightly with a
small amount of heavy-duty grease for load-bearing bolts (part no.
71401) and distribute the grease with a cloth so that it forms a thin
layer.
Refrigerated Centrifuge Sigma 2-16KL
8 Maintenance and service
53 / 82