Pos : 156 /100 Sigma/100 BA Z entrifugen Sigma (Standardm odul e)/060 Betrieb_Spi nc ontr ol L+S/Spinc ontrol L (Z ent6) /060- 0030-0030- 0010 Programm speic her n @ 27\m od_1405319309781_68.docx @ 193025 @ 4 @ 1
→
6.3.3.1
Saving a program
NOTE
NOTE
Pos : 157 /010 Univ ers almodul e/
Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 158 /100 Sigma/100 BA Z entrifugen Sigma (Standardm odul e)/060 Betrieb_Spi nc ontr ol L+S/Spinc ontrol L (Z ent6) /060- 0030-0030- 0020 Programm laden @ 27\mod_1405319312619_68.doc x @ 193053 @ 4 @ 1
6.3.3.2
Loading a program
Pos : 159 /010 Univ ers almodul e/
Seitenwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1
Version 08/2012, Rev. 1.24 of 28/04/2022 • sb
Translation of the original operating manual, part no. 0700702
This function is only available if the centrifuge is at a standstill.
•
Enter the parameters that are to be included in the program.
•
Select the menu item "save Prog" and confirm the selection. The
indication "save" flashes in the activated status.
•
Select a random storage location from the program selection list. Free
storage locations are indicated by a flashing display. Any storage
locations that are already occupied will be overwritten during the saving
process.
•
Save the program in the desired location.
The program is now saved.
The functions "Auto Open", "Buzzer", "ProgLock", and "Cycles" cannot be
entered as part of a program.
Fig. 26: Saving a program
•
Select the menu item "run Prog" and confirm the selection. The
indication "run" flashes in the activated status.
•
Select the desired program and confirm the selection by pressing the
function knob.
The program is now loaded.
Fig. 27: Loading a program
Refrigerated Centrifuge Sigma 2-16KL
6 Using the centrifuge
47 / 82