3.6.7

Service life

Pos : 57 /100 Sigma/100 BA Zentrifugen Sigma (Standar dmodule)/030 Sic her heit/030-0060-0064 Lebensdauer M asc hRL - T ext @ 26\mod_1405319032612_68.doc x @ 192143 @ @ 1
WARNING
NOTE
Pos : 58 /010 U nivers almodule/
Seitenw ec hsel @ 0\m od_1202116244312_0.docx @ 105 @ @ 1
Version 08/2012, Rev. 1.24 of 28/04/2022 • sb
Translation of the original operating manual, part no. 0700702
The rotors and accessories have a limited service life.
Perform regular checks (at least once per month) for safety reasons!
Pay special attention to changes, such as corrosion, cracks, material
abrasion, etc.
After 10 years, they must be inspected by the manufacturer.
After 50,000 cycles, the rotor must be scrapped for reasons of safety.
If other data concerning the service life are engraved on the rotor or
bucket, these data shall apply accordingly. For example, a bucket with
the engraving "max. cycles = 10,000" has a service life of 10,000
cycles, and a rotor with the engraving "Exp. date 01/27" must be
scrapped in January 2027 at the latest (see figure).
If a specification concerning the maximum number of cycles and a
specification concerning the service life (i.e. a date) are provided, the
specification that occurs first shall apply.
Fig. 4: Different service life – engraving on the bucket/rotor
Refer to the table of the service life of rotors and accessories (see
chapter 11.4 - " Table of the service life of rotors and accessories ")!