Emergency lid release

WARNING
Version 07/2012, Rev. 1.22 of 02/06/2021 • sb
Translation of the original operating manual, part no. 0701302
In the event of a power failure, it is possible to manually open the centrifuge
lid.
Switch off the mains power switch and disconnect the power cord from
the socket.
Remove the plug (see figure, item 1) from the opening on the left side of
the control panel, e.g. with a screwdriver.
Fig. 32: Position of the opening for the emergency lid release
Insert the supplied hexagon socket key horizontally into the hole. The
key will be guided through a funnel-shaped tube to the shaft of the lid
lock motor.
Fig. 33: The emergency lid release key must be inserted horizontally
(similar to illustration)
Unlock the motorised lid lock with the supplied hexagon socket key by
turning it anti-clockwise.
Then, reinsert the plug.
Do not unlock or open the lid unless the rotor is at a standstill.
If the lid is opened via the emergency lid release system during a centrifuge
run, the centrifuge will be switched off immediately and decelerate in an
unbraked manner.