7 Malfunctions and error correction
Pos: 160 /10 0 Sigm a/11 0 BA Z entrif uge n (Pr ojekt e)/6 -16S _6- 16HS/07 0 Stö run gen und Fe hlers uche /070 -00 10- 001 0 Not ent riegel ung 6- 16S @ 1 3\m od_ 134 035 984 7590 _68 .docx @ 6 773 9 @ 2 2 @ 1
7.1.1

Emergency lid release

WARNING
Pos: 161 /01 0 Univ ersal mod ule/
Sei tenwe chsel @ 0\ mod _12 021 1624 431 2_0 .docx @ 1 05 @ @ 1
50
In the event of a power failure, it is possible to manually open the
centrifuge lid.
Switch off the mains power switch and disconnect the power cord from
the socket.
Remove the plug (see figure, item 1) from the opening on the control
panel, e.g. with a screwdriver.
Fig. 26: Position of the opening for the emergency lid release
Insert the supplied hexagon socket key horizontally into the hole. The
key will be guided through a funnel-shaped tube to the shaft of the lid
lock motor.
Fig. 27: The emergency lid release key must be inserted horizontally
Unlock the motorised lid lock by turning it anti-clockwise.
Then, reinsert the plug.
Do not unlock or open the lid unless the rotor is at a standstill.
If the lid is opened via the emergency lid release system during a centrifuge
run, the centrifuge will be switched off immediately and decelerate in an
unbraked manner.
Version 06/2012, Rev. 1.0 • sb-kue
Translation of the original operating manual