Emergency Lid Release - Sigma 3-16L Operating Manual

Hide thumbs Also See for 3-16L:
Table of Contents

Advertisement

Laboratory Centrifuge Sigma 3-16L
Laboratory Centrifuge Sigma 3-16L IVD
7 Malfunctions and error correction
Pos : 166 /100 Sigma/110 BA Z entrifugen Sigma (Proj ekte)/3-16L/070 Störungen und F ehl ersuche/070- 0010- 0010 N otentriegel ung 3-16L @ 29\m od_1405321641164_68.docx @ 197281 @ 3 @ 1
7.1.1

Emergency lid release

WARNING
Pos : 167 /010 Univ ers almodul e/
Seitenwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1
50 / 84
In the event of a power failure, it is possible to manually open the centrifuge
lid.
Switch off the mains power switch and disconnect the power cord from
the socket.
Remove the stopper (see figure, item 1) from the opening at the side
panel, e.g. with a screwdriver.
Fig. 31: Position of the opening for the emergency lid release
First, pull the red string to release the mechanism and then unlock the
lid lock by pulling the white string.
Then, reinsert the plug and quit the error message by switching the
centrifuge off/on. Approximately 10 seconds after quitting the lid can be
closed again.
Do not unlock or open the lid unless the rotor is at a standstill.
If the lid is opened via the emergency lid release system during a centrifuge
run, the centrifuge will be switched off immediately and decelerate in an
unbraked manner.
Version 03/2014, Rev. 2.22 of 24/10/2023 • sb
Original operating manual, part no. 0700902