7 Malfunctions and error correction
Pos : 166 /100 Sigma/110 BA Z entrifugen Sigma (Proj ekte)/3-16L/070 Störungen und F ehl ersuche/070- 0010- 0010 N otentriegel ung 3-16L @ 29\m od_1405321641164_68.docx @ 197281 @ 3 @ 1
7.1.1
Emergency lid release
WARNING
Pos : 167 /010 Univ ers almodul e/
Seitenwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1
50 / 84
In the event of a power failure, it is possible to manually open the centrifuge
lid.
•
Switch off the mains power switch and disconnect the power cord from
the socket.
•
Remove the stopper (see figure, item 1) from the opening at the side
panel, e.g. with a screwdriver.
Fig. 31: Position of the opening for the emergency lid release
•
First, pull the red string to release the mechanism and then unlock the
lid lock by pulling the white string.
•
Then, reinsert the plug and quit the error message by switching the
centrifuge off/on. Approximately 10 seconds after quitting the lid can be
closed again.
Do not unlock or open the lid unless the rotor is at a standstill.
If the lid is opened via the emergency lid release system during a centrifuge
run, the centrifuge will be switched off immediately and decelerate in an
unbraked manner.
Version 03/2014, Rev. 2.22 of 24/10/2023 • sb