Refrigerated Centrifuge for Robot Use Sigma 4-5KRL
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD
7 Malfunctions and error correction
7.1.1

Emergency lid release

WARNING
52 / 113
In the event of a power failure, it is possible to manually open the centrifuge
lid.
Switch off the mains power switch and disconnect the power cord from
the socket.
Remove the stopper (see figure, item 1) from the opening on the right
side panel, e.g. with a screwdriver.
Fig. 31: Position of the opening for the emergency lid release
The stopper is connected with a string. Pull the string in order to release
the lid.
Then, reinsert the stopper into the opening.
Do not unlock or open the lid unless the rotor is at a standstill.
If the lid is opened via the emergency lid release system during a centrifuge
run, the centrifuge will be switched off immediately and decelerate in an
unbraked manner.
Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb
Translation of the original operating manual,